[THE VERIFICATION OF ANALYTICAL CHARACTERISTICS OF THREE MODELS OF GLUCOMETERS].

Autor: Timofeev AV, Khaibulina ET, Mamonov RA, Gorst KA
Jazyk: ruština
Zdroj: Klinicheskaia laboratornaia diagnostika [Klin Lab Diagn] 2016 Jan; Vol. 61 (1), pp. 39-45.
Abstrakt: The individual portable systems of control of glucose level in blood commonly known as glucometers permit to patients with diabetes mellitus to independently correct pharmaceutical therapy. The effectiveness of this correction depends on accuracy of control of glucose level. The evaluation was implemented concerning minimal admissible accuracy and clinical accuracy of control of glucose level of devices Contour TC, Satellite Express and One Touch Select according standards expounded in GOST 15197-2011 and international standard ISO 15197-2013. It is demonstrated that Contour TC and One Touch Select meet the requirements of these standards in part of accuracy while Satellite Express does not.