Autor: |
Savin DD; Department of Orthopedic Surgery, University of Illinois, Chicago, IL. ddsavin@gmail.com., Lee S, Bohnenkamp FC, Pastor A, Garapati R, Goldberg BA |
Jazyk: |
angličtina |
Zdroj: |
American journal of orthopedics (Belle Mead, N.J.) [Am J Orthop (Belle Mead NJ)] 2016 Mar-Apr; Vol. 45 (3), pp. E114-8. |
Abstrakt: |
In locking plate osteosynthesis, proper surgical technique is crucial in reducing potential pitfalls, and use of a torque limiter makes it possible to control insertion torque. We conducted a study of the ways in which different techniques can alter the accuracy of torque limiters. We tested 22 torque limiters (1.5 Nm) for accuracy using hand and power tools under different rotational scenarios: hand power at low and high velocity and drill power at low and high velocity. We recorded the maximum torque reached after each torque-limiting event. Use of torque limiters under hand power at low velocity and high velocity resulted in significantly (P < .0001) different mean (SD) measurements: 1.49 (0.15) Nm and 3.73 (0.79) Nm. Use under drill power at controlled low velocity and at high velocity also resulted in significantly (P < .0001) different mean (SD) measurements: 1.47 (0.14) Nm and 5.37 (0.90) Nm. Maximum single measurement obtained was 9.0 Nm using drill power at high velocity. Locking screw insertion with improper technique may result in higher than expected torque and subsequent complications. For torque limiters, the most reliable technique involves hand power at slow velocity or drill power with careful control of insertion speed until 1 torque-limiting event occurs. |
Databáze: |
MEDLINE |
Externí odkaz: |
|