Application of a point of care creatinine device for trend monitoring in kidney transplant patients: fit for purpose?

Autor: van Lint CL, van der Boog PJ, Romijn FP, Schenk PW, van Dijk S, Rövekamp TJ, Kessler A, Siekmann L, Rabelink TJ, Cobbaert CM
Jazyk: angličtina
Zdroj: Clinical chemistry and laboratory medicine [Clin Chem Lab Med] 2015 Sep 01; Vol. 53 (10), pp. 1547-56.
DOI: 10.1515/cclm-2014-0932
Abstrakt: Background: The StatSensor® Xpress-i™, a point-of-care system for blood creatinine measurement, offers patients the possibility of self-monitoring creatinine. In this study, the analytical performance of the StatSensor® for both detecting current renal function and monitoring renal (dys)function in kidney transplant patients was examined.
Methods: Accuracy of the StatSensor® with capillary and venous whole blood was evaluated and compared to an isotopic dilution mass spectrometry (IDMS)-traceable enzymatic creatinine test in venous serum (n=138). Twenty Li-heparin samples were compared to the IDMS reference method performed by a Joint Committee for Traceability in Laboratory Medicine (JCTLM)-listed reference laboratory (RfB, Bonn, Germany). To evaluate StatSensor®'s suitability to monitor kidney function, both venous and capillary samples were obtained in 20 hospitalized transplantation patients. Venous samples were analyzed with an IDMS-traceable enzymatic test, capillary samples were measured using the StatSensor®. For all 2-day intervals, percentage change in creatinine was compared between both methods.
Results: The StatSensor® did not meet total allowable error criterion of 6.9%. Average overall CVa for the StatSensor® was 10.4% and 5.2% for capillary and venous whole blood results, respectively. Overall CVa for the central laboratory serum creatinine method was <1.5%. For monitoring renal (dys)function, total agreement of the StatSensor® with an IDMS-traceable enzymatic test was 68% using a 10% Δ change. No significant differences were found between the changes observed by both methods.
Conclusions: Capillary blood testing with the StatSensor® is not advisable for determining current renal function with a single creatinine measurement in kidney transplant patients, mainly due to excessive analytical imprecision. However, our results suggest that capillary blood testing with the StatSensor® can be used for daily trend monitoring of kidney function after renal transplantation.
Databáze: MEDLINE