Validation of cone-beam computed tomography and magnetic resonance imaging of the porcine spine: a comparative study with multidetector computed tomography and anatomical specimens.

Autor: de Freitas RM; Radiology Department, Instituto de Radiologia-InRad, Faculdade de Medicina da Universidade de São Paulo, Av. Dr. Enéas de Carvalho Aguiar, s/n(o) - Rua 1 - Cerqueira César, CEP: 05403-900, São Paulo, SP, Brazil; Radiology Unit, Instituto do Câncer do Estado de São Paulo-ICESP, Avenida Doutor Arnaldo, 251, Cerqueira César, 01246-000, São Paulo, SP, Brazil. Electronic address: ricardomcfreitas@gmail.com., Andrade CS; Radiology Department, Instituto de Radiologia-InRad, Faculdade de Medicina da Universidade de São Paulo, Av. Dr. Enéas de Carvalho Aguiar, s/n(o) - Rua 1 - Cerqueira César, CEP: 05403-900, São Paulo, SP, Brazil., Caldas JG; Radiology Department, Instituto de Radiologia-InRad, Faculdade de Medicina da Universidade de São Paulo, Av. Dr. Enéas de Carvalho Aguiar, s/n(o) - Rua 1 - Cerqueira César, CEP: 05403-900, São Paulo, SP, Brazil., Kanas AF; Radiology Department, Instituto de Radiologia-InRad, Faculdade de Medicina da Universidade de São Paulo, Av. Dr. Enéas de Carvalho Aguiar, s/n(o) - Rua 1 - Cerqueira César, CEP: 05403-900, São Paulo, SP, Brazil., Cabral RH; Anatomy Department, University of São Paulo Institute of Biomedical Sciences, Av. Prof. Lineu Prestes, 2415 - Cidade Universitária, Butantã, 05508-900, São Paulo, SP, Brazil., Tsunemi MH; Biostatistics Department, Universidade Estadual Paulista Júlio de Mesquita Filho Biosciences Institute, Dist. Rubião Jr, 18618-970, Botucatu, SP, Brazil., Rodríguez HJ; Magnetic Resonance Imaging Laboratory, Instituto de Física da Universidade de São Paulo, Rua do Matão, Travessa R, 187, Cidade Universitária, 05508-090, São Paulo, Brazil., Rabbani SR; Magnetic Resonance Imaging Laboratory, Instituto de Física da Universidade de São Paulo, Rua do Matão, Travessa R, 187, Cidade Universitária, 05508-090, São Paulo, Brazil.
Jazyk: angličtina
Zdroj: The spine journal : official journal of the North American Spine Society [Spine J] 2015 May 01; Vol. 15 (5), pp. 992-9. Date of Electronic Publication: 2015 Jan 19.
DOI: 10.1016/j.spinee.2015.01.017
Abstrakt: Background Context: New spinal interventions or implants have been tested on ex vivo or in vivo porcine spines, as they are readily available and have been accepted as a comparable model to human cadaver spines. Imaging-guided interventional procedures of the spine are mostly based on fluoroscopy or, still, on multidetector computed tomography (MDCT). Cone-beam computed tomography (CBCT) and magnetic resonance imaging (MRI) are also available methods to guide interventional procedures. Although some MDCT data from porcine spines are available in the literature, validation of the measurements on CBCT and MRI is lacking.
Purpose: To describe and compare the anatomical measurements accomplished with MDCT, CBCT, and MRI of lumbar porcine spines to determine if CBCT and MRI are also useful methods for experimental studies.
Study Design: An experimental descriptive-comparative study.
Methods: Sixteen anatomical measurements of an individual vertebra from six lumbar porcine spines (n=36 vertebrae) were compared with their MDCT, CBCT, and MRI equivalents. Comparisons were made for the absolute values of the parameters.
Results: Similarities were found in all imaging methods. Significant correlation (p<.05) was observed with all variables except those that included cartilaginous tissue from the end plates when the anatomical study was compared with the imaging methods.
Conclusions: The CBCT and MRI provided imaging measurements of the lumbar porcine spines that were similar to the anatomical and MDCT data, and they can be useful for specific experimental research studies.
(Copyright © 2015 Elsevier Inc. All rights reserved.)
Databáze: MEDLINE