Inter-machine reliability of the Biodex and Cybex isokinetic dynamometers for knee flexor/extensor isometric, concentric and eccentric tests.

Autor: de Araujo Ribeiro Alvares JB; Exercise Research Laboratory (LAPEX), Federal University of Rio Grande do Sul (UFRGS), Rua Felizardo, 750, Bairro Jardim Botânico, CEP: 90690-200 Porto Alegre, RS, Brazil., Rodrigues R; Exercise Research Laboratory (LAPEX), Federal University of Rio Grande do Sul (UFRGS), Rua Felizardo, 750, Bairro Jardim Botânico, CEP: 90690-200 Porto Alegre, RS, Brazil., de Azevedo Franke R; Exercise Research Laboratory (LAPEX), Federal University of Rio Grande do Sul (UFRGS), Rua Felizardo, 750, Bairro Jardim Botânico, CEP: 90690-200 Porto Alegre, RS, Brazil., da Silva BG; Exercise Research Laboratory (LAPEX), Federal University of Rio Grande do Sul (UFRGS), Rua Felizardo, 750, Bairro Jardim Botânico, CEP: 90690-200 Porto Alegre, RS, Brazil., Pinto RS; Exercise Research Laboratory (LAPEX), Federal University of Rio Grande do Sul (UFRGS), Rua Felizardo, 750, Bairro Jardim Botânico, CEP: 90690-200 Porto Alegre, RS, Brazil., Vaz MA; Exercise Research Laboratory (LAPEX), Federal University of Rio Grande do Sul (UFRGS), Rua Felizardo, 750, Bairro Jardim Botânico, CEP: 90690-200 Porto Alegre, RS, Brazil., Baroni BM; Exercise Research Laboratory (LAPEX), Federal University of Rio Grande do Sul (UFRGS), Rua Felizardo, 750, Bairro Jardim Botânico, CEP: 90690-200 Porto Alegre, RS, Brazil. Electronic address: bmbaroni@yahoo.com.br.
Jazyk: angličtina
Zdroj: Physical therapy in sport : official journal of the Association of Chartered Physiotherapists in Sports Medicine [Phys Ther Sport] 2015 Feb; Vol. 16 (1), pp. 59-65. Date of Electronic Publication: 2014 May 15.
DOI: 10.1016/j.ptsp.2014.04.004
Abstrakt: Objectives: To assess the inter-machine reliability of the Biodex System 3 Pro and Cybex Humac Norm Model 770 dynamometers for knee extensor and knee flexor peak torque measurements in isometric, concentric and eccentric tests.
Design: Randomized/crossover.
Setting: Exercise Research Laboratory, Federal University of Rio Grande do Sul (Brazil).
Participants: 25 healthy male subjects.
Main Outcome Measures: Isometric, concentric and eccentric knee extensor and knee flexor peak torques recorded in the same test procedure performed on both isokinetic dynamometers. One-way ANOVA, intraclass correlation coefficient (ICC), standard error of measurement (SEM) and coefficient of variation (CV) were used to verify significant differences, relative and absolute reliability between devices.
Results: No significant differences were found between tests performed on Biodex and Cybex (p > 0.05). ICC values indicated a high to very high reproducibility for isometric, concentric and eccentric peak torques (0.88-0.92), and moderate to high reliability for agonist-antagonist strength ratios (0.62-0.73). Peak torque did not show great difference between dynamometers for SEM (3.72-11.27 Nm) and CV (5.27-7.77%). Strength ratios presented CV values of 8.57-10.72%.
Conclusion: Maximal knee extensor and knee flexor tests performed in isometric (60° of knee flexion), concentric and eccentric modes at 60°/s in Biodex and Cybex dynamometers present similar values.
(Copyright © 2014 Elsevier Ltd. All rights reserved.)