Effect of geometrical parameters on the performance of longitudinal functionally graded femoral prostheses.

Autor: Oshkour AA; Department of Mechanical Engineering, Faculty of Engineering, University of Malaya, Kuala Lumpur, Malaysia., Talebi H, Seyed Shirazi SF, Yau YH, Tarlochan F, Abu Osman NA
Jazyk: angličtina
Zdroj: Artificial organs [Artif Organs] 2015 Feb; Vol. 39 (2), pp. 156-64. Date of Electronic Publication: 2014 May 20.
DOI: 10.1111/aor.12315
Abstrakt: This study aimed to assess the performance of different longitudinal functionally graded femoral prostheses. This study was also designed to develop an appropriate prosthetic geometric design for longitudinal functionally graded materials. Three-dimensional models of the femur and prostheses were developed and analyzed. The elastic modulus of these prostheses in the sagittal plane was adjusted along a gradient direction from the distal end to the proximal end. Furthermore, these prostheses were composed of titanium alloy and hydroxyapatite. Results revealed that strain energy, interface stress, and developed stress in the femoral prosthesis and the bone were influenced by prosthetic geometry and gradient index. In all of the prostheses with different geometries, strain energy increased as gradient index increased. Interface stress and developed stress decreased. The minimum principal stress and the maximum principal stress of the bone slightly increased as gradient index increased. Hence, the combination of the femoral prosthetic geometry and functionally graded materials can be employed to decrease stress shielding. Such a combination can also be utilized to achieve equilibrium in terms of the stress applied on the implanted femur constituents; thus, the lifespan of total hip replacement can be prolonged.
(Copyright © 2014 International Center for Artificial Organs and Transplantation and Wiley Periodicals, Inc.)
Databáze: MEDLINE