Procedure for microbial identification based on Matrix-Assisted Laser Desorption/Ionization-Time of Flight Mass Spectrometry from screening-positive urine samples.
Autor: | March Rosselló GA; Service of Microbiology, University Clinic Hospital of Valladolid, Valladolid, Spain., Gutiérrez Rodríguez MP, de Lejarazu Leonardo RO, Orduña Domingo A, Bratos Pérez MA |
---|---|
Jazyk: | angličtina |
Zdroj: | APMIS : acta pathologica, microbiologica, et immunologica Scandinavica [APMIS] 2014 Sep; Vol. 122 (9), pp. 790-5. Date of Electronic Publication: 2013 Dec 10. |
DOI: | 10.1111/apm.12208 |
Abstrakt: | Matrix-Assisted Laser Desorption/Ionization-Time of Flight Mass Spectrometry is a widely used proteomic technique in clinical microbiology laboratories, and enables microbial identification directly from clinical samples. This study seeks to establish a protocol for bacterial identification from monomicrobial urine samples that have tested positive in the screening with Sysmex UF-1000i (Sysmex Corporation, Kobe, Japan). Sysmex UF-1000i counts ≥1 × 10(7) bacteria/mL indicate a sufficient bacterial concentration to allow direct identification from urine, with 87.5% sensitivity. Microbial identification from urine with Sysmex UF-1000i counts between 1 × 10(5) and 1 × 10(7) bacteria/ml requires preincubation to obtain the adequate amount of bacteria needed for analysis, and 91.7% sensitivity thus being achieved. (© 2013 APMIS. Published by John Wiley & Sons Ltd.) |
Databáze: | MEDLINE |
Externí odkaz: | |
Nepřihlášeným uživatelům se plný text nezobrazuje | K zobrazení výsledku je třeba se přihlásit. |