Image quality, threshold contrast and mean glandular dose in CR mammography.

Autor: Jakubiak RR; Academic Department of Physics, Federal University of Technology-Paraná, Av. Sete de Setembro, 3165, Centro, 80230-010, Curitiba/PR, Brazil. requi@utfpr.edu.br, Gamba HR, Neves EB, Peixoto JE
Jazyk: angličtina
Zdroj: Physics in medicine and biology [Phys Med Biol] 2013 Sep 21; Vol. 58 (18), pp. 6565-83. Date of Electronic Publication: 2013 Sep 04.
DOI: 10.1088/0031-9155/58/18/6565
Abstrakt: In many countries, computed radiography (CR) systems represent the majority of equipment used in digital mammography. This study presents a method for optimizing image quality and dose in CR mammography of patients with breast thicknesses between 45 and 75 mm. Initially, clinical images of 67 patients (group 1) were analyzed by three experienced radiologists, reporting about anatomical structures, noise and contrast in low and high pixel value areas, and image sharpness and contrast. Exposure parameters (kV, mAs and target/filter combination) used in the examinations of these patients were reproduced to determine the contrast-to-noise ratio (CNR) and mean glandular dose (MGD). The parameters were also used to radiograph a CDMAM (version 3.4) phantom (Artinis Medical Systems, The Netherlands) for image threshold contrast evaluation. After that, different breast thicknesses were simulated with polymethylmethacrylate layers and various sets of exposure parameters were used in order to determine optimal radiographic parameters. For each simulated breast thickness, optimal beam quality was defined as giving a target CNR to reach the threshold contrast of CDMAM images for acceptable MGD. These results were used for adjustments in the automatic exposure control (AEC) by the maintenance team. Using optimized exposure parameters, clinical images of 63 patients (group 2) were evaluated as described above. Threshold contrast, CNR and MGD for such exposure parameters were also determined. Results showed that the proposed optimization method was effective for all breast thicknesses studied in phantoms. The best result was found for breasts of 75 mm. While in group 1 there was no detection of the 0.1 mm critical diameter detail with threshold contrast below 23%, after the optimization, detection occurred in 47.6% of the images. There was also an average MGD reduction of 7.5%. The clinical image quality criteria were attended in 91.7% for all breast thicknesses evaluated in both patient groups. Finally, this study also concluded that the use of the AEC of the x-ray unit based on the constant dose to the detector may bring some difficulties to CR systems to operate under optimal conditions. More studies must be performed, so that the compatibility between systems and optimization methodologies can be evaluated, as well as this optimization method. Most methods are developed for phantoms, so comparative studies including clinical images must be developed.
Databáze: MEDLINE