A technical assessment of pulse wave velocity algorithms applied to non-invasive arterial waveforms.

Autor: Gaddum NR; Division of Imaging Sciences and Biomedical Engineering, St. Thomas' Hospital, King's College London, London, UK, nickgaddum@gmail.com., Alastruey J, Beerbaum P, Chowienczyk P, Schaeffter T
Jazyk: angličtina
Zdroj: Annals of biomedical engineering [Ann Biomed Eng] 2013 Dec; Vol. 41 (12), pp. 2617-29. Date of Electronic Publication: 2013 Jul 02.
DOI: 10.1007/s10439-013-0854-y
Abstrakt: Non-invasive assessment of arterial stiffness through pulse wave velocity (PWV) analysis is becoming common clinical practice. However, the effects of measurement noise, temporal resolution and similarity of the two waveforms used for PWV calculation upon accuracy and variability are unknown. We studied these effects upon PWV estimates given by foot-to-foot, least squared difference, and cross-correlation algorithms. We assessed accuracy using numerically generated blood pressure and flow waveforms for which the theoretical PWV was known to compare with the algorithm estimates. We assessed variability using clinical measurements in 28 human subjects. Wave shape similarity was quantified using a cross correlation-coefficient (CCCoefficient), which decreases with increasing distance between waveform measurements sites. Based on our results, we propose the following criteria to identify the most accurate and least variable algorithm given the noise, resolution and CCCoefficient of the measured waveforms. (1) Use foot-to-foot when the noise-to-signal ratio ≤10%, and/or temporal resolution ≥100 Hz. Otherwise (2) use a least squares differencing method applied to the systolic upstroke.
Databáze: MEDLINE