Implications of complex anatomical junctions on conductance catheter measurements of coronary arteries.

Autor: Choi HW; Department of Biomedical Engineering, Indiana University-Purdue University Indianapolis, Indianapolis, Indiana 46202, USA., Zhang ZD, Farren ND, Kassab GS
Jazyk: angličtina
Zdroj: Journal of applied physiology (Bethesda, Md. : 1985) [J Appl Physiol (1985)] 2013 Mar 01; Vol. 114 (5), pp. 656-64. Date of Electronic Publication: 2013 Jan 10.
DOI: 10.1152/japplphysiol.00987.2012
Abstrakt: In vivo, the position of the conductance catheter to measure vessel lumen cross-sectional area may vary depending on where the conductance catheter is deployed in the complex anatomical geometry of arteries, including branches, bifurcations, or curvatures. The objective here is to determine how such geometric variations affect the cross-sectional area (CSA) estimates obtained using the cylindrical model. Computer simulations and in vitro and in vivo experiments were used to assess how the electric field and associated CSA measurement accuracy are affected by three typical in vivo conditions: 1) a vessel with abrupt change in lumen diameter (e.g., transition from aorta to coronary ostia); 2) a vessel with a T-bifurcation or a Y-bifurcation; and 3) a vessel curvature, such as in the right coronary artery, aorta, or pulmonary artery. The error in diameter from simulation results was shown to be relatively small (<7%), unless the detection electrodes were placed near the junction between two different lumen diameters or at a bifurcation junction. Furthermore, the present findings show that the effect of misaligned catheter-vessel geometrical configuration and vessel curvature on measurement accuracy is negligible. Collectively, the findings support the accuracy of the conductance method for sizing blood vessels, despite the geometric complexities of the cardiovascular system, as long as the detection electrodes are not placed at a large discontinuity in diameter or at bifurcation junctions.
Databáze: MEDLINE