Impact of the analytical blank in the uncertainty evaluation of the copper content in waters by flame atomic absorption spectrometry.

Autor: de Oliveira EC; PETROBRAS TRANSPORTE S.A., Gerência Geral de Engenharia, Av. Pres. Vargas, 328, 20091-060 Rio de Janeiro, RJ, Brazil. elciooliveira@petrobras.com.br, Monteiro MI, Pontes FV, de Almeida MD, Carneiro MC, da Silva LI, Alcover Neto A
Jazyk: angličtina
Zdroj: Journal of AOAC International [J AOAC Int] 2012 Mar-Apr; Vol. 95 (2), pp. 560-6.
DOI: 10.5740/jaoacint.11-018
Abstrakt: Chemical analysts use analytical blanks in their analyses, but seldom is this source of uncertainty evaluated. Generally, there is great confusion. Although the numerical value of the blank, in some situations, can be negligible, its source of uncertainty cannot be. This article discusses the uncertainty contribution of the analytical blank using a numerical example of the copper content in waters by flame atomic absorption spectrometry. The results indicate that the uncertainties of the analytical blank can contribute up to 50% when the blank sample is considered in this analysis, confirming its high impact. This effect can be primarily observed where the analyte concentration approaches the lower range of the analytical curve. Even so, the blank is not always computed. Therefore, the relevance of the analytical blank can be confirmed by uncertainty evaluation.
Databáze: MEDLINE