Semi-automatic version of the potentiometric titration method for characterization of uranium compounds.

Autor: Cristiano BF; Comissão Nacional de Energia Nuclear (CNEN), Instituto de Radioproteção e Dosimetria (IRD), Avenida Salvador Allende, s/n Recreio dos Bandeirantes 22780-160, Rio de Janeiro, RJ, Brasil. barbara@ird.gov.br, Delgado JU, da Silva JW, de Barros PD, de Araújo RM, Dias FC, Lopes RT
Jazyk: angličtina
Zdroj: Applied radiation and isotopes : including data, instrumentation and methods for use in agriculture, industry and medicine [Appl Radiat Isot] 2012 Sep; Vol. 70 (9), pp. 1952-4. Date of Electronic Publication: 2012 Feb 27.
DOI: 10.1016/j.apradiso.2012.02.051
Abstrakt: The potentiometric titration method was used for characterization of uranium compounds to be applied in intercomparison programs. The method is applied with traceability assured using a potassium dichromate primary standard. A semi-automatic version was developed to reduce the analysis time and the operator variation. The standard uncertainty in determining the total concentration of uranium was around 0.01%, which is suitable for uranium characterization and compatible with those obtained by manual techniques.