Commissioning of a displacement-controlled knee wear simulator and exploration of some issues related to the lubricant.

Autor: Brandt JM; Concordia Joint Replacement Group, Concordia Hip & Knee Institute, Winnipeg, Canada. jbrandt@cjrg.ca, Charron KD, Zhao L, MacDonald SJ, Medley JB
Jazyk: angličtina
Zdroj: Proceedings of the Institution of Mechanical Engineers. Part H, Journal of engineering in medicine [Proc Inst Mech Eng H] 2011 Aug; Vol. 225 (8), pp. 736-52.
DOI: 10.1177/0954411911406061
Abstrakt: A six-station displacement-controlled knee simulator with separately controlled left (L) and right (R) banks (three wear implants per bank) was commissioned for a total of three million cycles (Mc) following ISO 14243-3. A commissioning protocol was applied to compare the polyethylene wear among the six wear stations by exchanging the implants between wear stations. Changes in lubricant characteristics during wear testing, such as polypeptide degradation, low-molecular-weight polypeptide concentration, and possible microbial contamination were also assessed. The total mean wear rate for the implants was 23.60 +/- 1.96 mm3/Mc and this was of a similar magnitude to the mean wear rate for the same implant tested under similar conditions by DePuy Orthopaedics Inc. (Warsaw, IN). Repeated run-in wear was observed when the implants were exchanged between wear stations, suggesting that implants should be subjected to the same wear station throughout the duration of a wear test. The total polypeptide degradation for the implants measured 30.53 +/- 3.96 percent; the low-molecular-weight polypeptide concentration of the "used" lubricant for implants (0.131 +/- 0.012 g/L) was 3.3 times greater than the mean polypeptide concentration of the fresh, "unused" lubricant (0.039 +/- 0.004 g/L). This increase in low-molecular weight polypeptide concentration was suggested to be attributable to protein shear in the articulation of the implant, the circulation of the lubricant, and some proteolytic activity. Sodium azide was ineffective in maintaining a sterile environment for wear testing as a single, highly motile Gram-negative micro-organism was identified in the lubricant from wear tests.
Databáze: MEDLINE