Characterization of PEEK biomaterials using the small punch test.

Autor: Jaekel DJ; School of Biomedical Engineering, Science, and Health Systems, Drexel University, 3141 Chestnut Street Philadelphia, PA 19104, USA. djj25@drexel.edu, MacDonald DW, Kurtz SM
Jazyk: angličtina
Zdroj: Journal of the mechanical behavior of biomedical materials [J Mech Behav Biomed Mater] 2011 Oct; Vol. 4 (7), pp. 1275-82. Date of Electronic Publication: 2011 Apr 22.
DOI: 10.1016/j.jmbbm.2011.04.014
Abstrakt: The small punch test is widely used to characterize the ductility and fracture resistance in metals and ceramics, when only a small volume of material is available. This study was conducted to investigate the suitability of the small punch test for characterizing polyetheretherketone (PEEK) polymeric biomaterials for changes in material grade, crystallinity, and molding process. The small punch test reproducibly characterized the mechanical behavior of PEEK and was able to distinguish differences induced by molding process alterations and annealing. Peak load was most sensitive to changes in crystallinity, grade, molding process, and increased with increasing crystallinity, but decreased with the addition of image contrast materials. The ultimate displacement was negatively correlated with crystallinity. Molding process conditions had the greatest influence on metrics of the small punch test, when compared with the effects of annealing and the addition of a radiopacifier. The results of this study validate the small punch test as a repeatable method for measuring the mechanical behavior of PEEK biomaterials.
(Copyright © 2011 Elsevier Ltd. All rights reserved.)
Databáze: MEDLINE