Autor: |
Gemperline PJ; Department of Chemistry, East Carolina University, Greenville, North Carolina 27858., Cho J, Aldridge PK, Sekulic SS |
Jazyk: |
angličtina |
Zdroj: |
Analytical chemistry [Anal Chem] 1996 Sep 01; Vol. 68 (17), pp. 2913-5. |
DOI: |
10.1021/ac9604191 |
Abstrakt: |
Several years ago, we noted that spectra transformed by the piecewise direct standardization (PDS) method may contain discontinuities. Having noticed that the problem was a recurring one, we studied it and recently diagnosed its source. Our investigations suggest that this problem also occurs in applications of window factor analysis, evolving factor analysis, and any other procedure that uses piecewise principal component models. In this work, we report the source of the problem and illustrate it with one example. A procedure is presented for eliminating the problem that is effective in PDS pattern recognition applications. Further work is needed to develop modified algorithms suitable for calibration applications. |
Databáze: |
MEDLINE |
Externí odkaz: |
|