Dosimetry of gamma chamber blood irradiator using Gafchromic EBT film.
Autor: | Vandana S; Radiological Physics and Advisory Division, Bhabha Atomic Research Centre, CT&CRS Building, Anushaktinagar, Mumbai 400094, India., Shaiju VS, Sharma SD, Mhatre S, Shinde S, Chourasiya G, Mayya YS |
---|---|
Jazyk: | angličtina |
Zdroj: | Applied radiation and isotopes : including data, instrumentation and methods for use in agriculture, industry and medicine [Appl Radiat Isot] 2011 Jan; Vol. 69 (1), pp. 130-5. Date of Electronic Publication: 2010 Sep 17. |
DOI: | 10.1016/j.apradiso.2010.08.018 |
Abstrakt: | Gafchromic films are increasingly being used for dosimetry in medical and industrial applications of ionizing radiation because of their favorable characteristics such as self developing in nature, easy to handle, dose rate independent response, insensitivity to normal room lights, high resolution and insensitivity to variation in the environment conditions. The dosimetry measurements using Gafchromic EBT film, Fricke dosimeter and thermoluminescence (TL) dosimeter were carried out on gamma blood irradiators to establish the suitability of Gafchromic EBT film in such applications. The dose rates determined by Gafchromic EBT film, Fricke dosimeter and TL dosimeter powder are found in good agreement to one another within the uncertainty of measurement. The two-dimensional dose distributions determined using Gafchromic EBT film reflects the expected dose distribution inside the sample compartment of the blood irradiator. The results of this study establish Gafchromic EBT film a suitable dosimeter for routine dosimetry on gamma blood irradiators. (Copyright © 2010 Elsevier Ltd. All rights reserved.) |
Databáze: | MEDLINE |
Externí odkaz: |