Autor: |
Ji CH; School of Applied Science, University of Science and Technology Beijing, Beijing 100083, China. 43489100@163.com, Li JQ, Huang WJ, Bao R, Guo R, Hu JJ |
Jazyk: |
čínština |
Zdroj: |
Guang pu xue yu guang pu fen xi = Guang pu [Guang Pu Xue Yu Guang Pu Fen Xi] 2010 May; Vol. 30 (5), pp. 1396-9. |
Abstrakt: |
A method for the determination of trace gold by inductively coupled plasma atomic emission spectrometry (ICP-AES) in mine sample was proposed. The instrument parameters were optimized and the media of solution and interferences were studied systemically. The results showed that in less than 10% (phi) aqua regia medium, the acids had no influences on the determination, and after removing common cations in mineral samples by using 10% hydrochloric acid(phi), the residual matrix elements had no effect on the determination of Au with ICP-AES. The silicon in samples had serious spectral interference to Au 208. 2 nm, and the spectral interference was corrected with interference coefficient using the 251.6 nm silicon line. With 10% HCl to remove the matrix and interference correction coefficient to eliminate the interference of silicon, trace gold in high-sulfur tailings samples was determined, the result was identical with the reference values by the method of enrichment with activated carbon-atomic absorption spectrometry. Under the experimental conditions, the detection limit for gold in mine sample was 0.10 g x t(-1). |
Databáze: |
MEDLINE |
Externí odkaz: |
|