Paper is devoted to the safety rules by surgical interventions on patients with permanent pacemaker. Data of 123 operated patients with intraoperative pacemaker instability were analyzed, possible premises of this complications were discussed. The use of monopolar electrocoagulation proved to be the most frequent of reason of pacemaker failure. Strong correlation between rhythm failure and duration of the single electrocautery exposure was revealed. |