Endoscopic ultrasound guided fine-needle aspiration core biopsy: comparison between an automatic biopsy device and two conventional needle systems.

Autor: Ardengh JC; Endoscopy Unit, Division of Gastroenterology, Universidade Federal de São Paulo, Brazil. jcelso@uol.com.br, Paulo GA, Nakao FS, Venco F, Santo GC, Geocze S
Jazyk: angličtina
Zdroj: Acta gastroenterologica Latinoamericana [Acta Gastroenterol Latinoam] 2008 Jun; Vol. 38 (2), pp. 105-15.
Abstrakt: Background: endoscopic ultrasound guided fine-needle aspiration (EUS-FNA) allows cytologic and/or histologic diagnosis of lesions within or adjacent to the gastrointestinal tract. However, the amount of tissue obtained with a regular 22 gauge needle is not always satisfactory. With the development of a needle XNA-10J-KB (Shot-Gun) that resembles the automatic liver biopsy needle, it is expected that significant samples be obtained more frequently (core biopsy), optimizing histological analysis.
Objective: to compare samples obtained with EUS-FNA using 3 different needle systems: GIP, NA-10J-1 and Shot-Gun.
Methods: 19 patients underwent EUS-FNA for diagnosis (5) or tumor staging (14). Mean age was 58.9 years (range 27-82), being 50% men. All patients were submitted to EUS-FNA with the 3 needle models. The Shot-Gun model was "shot" when its tip was near the target inside the lesion, followed by aspiration. Samples were submitted for cytologic and histologic examination.
Results: mean lesion size was 3.0 cm (range 0.8-5.5 cm). Final diagnoses were made after surgery or intra-operative biopsy: 13 pancreatic tumors (12 adenocarcinomas and 1 neuroendocrine tumor), 4 chronic pancreatitis, 1 acute pancreatitis, and 1 cholangiocarcinoma. Specimens adequate for cytologic diagnosis were obtained in 13/19 (68. 4%) patients using GIP model, in 14/19 (73.7%) with NA10J-1 model, and in 17/19 (89.5%) with ShotGun, model (p=0.039). Histologic analysis was possible in 10/19 (52.6%) patients using the GIP model, in 14/19 (73.7%) with NA10J-1, and in 17/19 (89.5%) with Shot-Gun, model (p=0.005). Adequate samples for cytologic or histologic assessment in 16/19 (84.2%) patients using the GIP model, in 17/19 (89.5%) with NA10J-1, and in 18/19 (94.7%) with Shot-Gun, model (p=0.223). In two cases biopsies were negative due to very hard tumors.
Conclusion: the Shot-Gun needle obtained better samples for histological diagnosis than NA10J-1 needle and GIP.
Databáze: MEDLINE