Measurement uncertainty assessment of the scanning laser acoustic microscope and application to canine skin and wound.

Autor: Steiger DL; Illinois Univ., Urbana, IL., O'Brien WR, Olerud JE, Riederer-Henderson MA, Odland GF
Jazyk: angličtina
Zdroj: IEEE transactions on ultrasonics, ferroelectrics, and frequency control [IEEE Trans Ultrason Ferroelectr Freq Control] 1988; Vol. 35 (6), pp. 741-8.
DOI: 10.1109/58.9331
Abstrakt: Accuracy and precision were experimentally evaluated for the scanning laser acoustic microscope (SLAM) by measurements on homogeneous liquids of known ultrasonic properties. Using aqueous solutions of bovine serum albumin, the attenuation coefficient accuracy and precision are +/-12% and +/-15%, respectively. Using Dow Corning 710, a silicone oil, the speed accuracy and precision are +/-2.9% and +/-0.4%, respectively. Precision was assessed using duplicate samples of canine skin and wound tissue. The estimated precisions in the measurement of the attenuation coefficient and speed were +/-16% and +/-1.7%, respectively.