Overestimation of aortic valve gradient measured by Doppler echocardiography in patients with aortic stenosis.
Autor: | Sakthi C; Department of Cardiology, Kaiser Permanente Medical Center, Los Angeles, California 90027, USA., Yee H, Kotlewski A |
---|---|
Jazyk: | angličtina |
Zdroj: | Catheterization and cardiovascular interventions : official journal of the Society for Cardiac Angiography & Interventions [Catheter Cardiovasc Interv] 2005 Jun; Vol. 65 (2), pp. 176-9. |
DOI: | 10.1002/ccd.20324 |
Abstrakt: | We describe a series of patients with suspected aortic stenosis who were noted to have a significant aortic valve gradient by Doppler echocardiography but only a minimal gradient during cardiac catheterization. Although the pressure recovery phenomenon partially explained this discrepancy, a difference still persisted, particularly in peak gradients. The overestimation occurred at much lower cardiac catheterization gradients than previously reported. Since many of our patients were noted to be hypertensive, this discrepancy might be more apparent in this patient population. (Copyright 2005 Wiley-Liss, Inc) |
Databáze: | MEDLINE |