J-tip spring guidewire entrapment by an inferior vena cava filter.

Autor: Vinces FY; Trauma and Critical Care Section, Department of Surgery, St Barnabas Hospital, Bronx, New York 10457, USA. vincesf@optonline.net, Robb TV, Alapati K, Brunson VW, Hutchison P, Verrier RJ, Kemp L, Fried G
Jazyk: angličtina
Zdroj: The Journal of the American Osteopathic Association [J Am Osteopath Assoc] 2004 Feb; Vol. 104 (2), pp. 87-9.
Abstrakt: The number of complications between guidewires and inferior vena cava filters is unknown and most likely underreported. Since 1993, at least 17 cases of central venous catheter guidewires entangled in inferior vena cava filters have been reported. The placement of both devices in the intensive care setting has increased the number of incidents in which a guidewire from a central venous catheter becomes entrapped in an inferior vena cava filter. The authors report a case in which entrapment of a guidewire occurred without causing displacement of the filter. In addition, a review of simple but useful recommendations to prevent and manage these complications is presented.