3D time of flight MR angiography: acquisition with small field of view and low phase encodes.

Autor: Chan JH; Department of Diagnostic Radiology, Tuen Mun Hospital, Tsing Chung Koon Road, Tuen Mun, NT, Hong Kong, People's Republic of China. jchm888@yahoo.com, Tsui EY, Kwong KL, Poon WL, Yuen MK, Leung JO, Lai WT, Yu CW, Wong KP
Jazyk: angličtina
Zdroj: Computerized medical imaging and graphics : the official journal of the Computerized Medical Imaging Society [Comput Med Imaging Graph] 2003 Jul-Aug; Vol. 27 (4), pp. 283-7.
DOI: 10.1016/s0895-6111(02)00092-7
Abstrakt: The aim of the study was to evaluate and compare the image quality of the 3D TOF MRA acquired with a small FOV and low phase encodes with those MR angiographic images acquired with standard pulse sequence parameters. Twenty patients who were referred to our institution for MR imaging of the brain and strictly satisfied the selection criteria were included in this study. Apart from the routine protocol for MR imaging of the brain, 3D TOF MRA of the circle of Willis with a small FOV and a standard FOV were performed. The image quality of all MRA was evaluated by two independent observers who were blind to the pulse sequence parameters. From the standard FOV MRA, 22.5, 12.5, and 5% of the patients were graded as mild, moderate, and severe stenosis of the internal carotid artery, respectively. On the contrary, no apparent stenosis was observed from the small FOV MRA with low phase encodes. Regarding the reduction in MR artifacts and acquisition time achieved with the small FOV 3D TOF MRA with low phase encodes, this might be a useful MR angiographic technique to be used in routine clinical practice.
Databáze: MEDLINE