Miniature analyzer Glucophot BP-"M", intended for express measurements of glucose concentrations in capillary blood, was tested under clinical conditions. The reference method was the hexokinase method for glucose measurements on a Spectrum 11 biochemical analyzer (Abbott); in addition, glucose measurements were carried out on a One Touch-11 glucometer (Johnson-Johnson). The results evidence that the reproducibility and correctness of glucose analysis by the Glucophot BP-"M" analyzer both in health and disease met the requirements to laboratory measurements performed by "dry" chemical methods and demonstrated its high analytical reliability. |