Does stent design affect probability of restenosis? A randomized trial comparing Multilink stents with GFX stents.

Autor: Yoshitomi Y; Division of Cardiology, Tohsei National Hospital, Shizuoka, Japan. ytommy@sage.ocn.ne.jp, Kojima S, Yano M, Sugi T, Matsumoto Y, Saotome M, Tanaka K, Endo M, Kuramochi M
Jazyk: angličtina
Zdroj: American heart journal [Am Heart J] 2001 Sep; Vol. 142 (3), pp. 445-51.
DOI: 10.1067/mhj.2001.117321
Abstrakt: Background: Experimental studies have revealed that stent configuration influences intimal hyperplasia. The purpose of this study was to evaluate clinical outcomes for 2 stent designs in a randomized trial with quantitative coronary angiography (QCA) and intravascular ultrasonography (IVUS).
Methods: We randomly assigned 100 patients with 107 lesions and symptomatic coronary artery disease to deployment of a Multilink stent (Advanced Cardiovascular Systems, Guidant, Santa Clara, Calif) or a GFX stent (Applied Vascular Engineering, Santa Rosa, Calif) with IVUS guidance. QCA and IVUS studies were performed before and after intervention and at follow-up (4.2 +/- 1.0 months).
Results: There were no significant differences in baseline characteristics and QCA and IVUS parameters before and after intervention between the 2 groups. However, minimal lumen diameter at follow-up was significantly larger in the Multilink group (2.46 +/- 0.59 vs 2.08 +/- 0.79 mm, P <.05). Maximal in-stent intimal hyperplasia was significantly larger in the GFX group (2.9 +/- 1.7 vs 1.8 +/- 1.2 mm(2), P <.01). The restenosis rate differed between the 2 groups (Multilink 4% vs GFX 26%, P =.003). In multiple stepwise logistic regression analysis, the only predictor that significantly correlated with restenosis was stent type (P <.01). The odds ratio for the GFX stent-treated vessels was 18.65 (95% confidence interval 2.10-165.45).
Conclusions: With deployment of the GFX stent, a thicker neointima develops within the stent. Stent configuration may affect clinical outcomes.