Autor: |
Frank WE; Illinois State Police, Research and Development Laboratory, Springfield 62702, USA., Llewellyn BE, Fish PA, Riech AK, Marcacci TL, Gandor DW, Parker D, Carter RR, Thibault SM |
Jazyk: |
angličtina |
Zdroj: |
Journal of forensic sciences [J Forensic Sci] 2001 May; Vol. 46 (3), pp. 642-6. |
Abstrakt: |
According to TWGDAM guideline 4.5 (1), prior to implementing a new DNA analysis procedure or an existing DNA analysis procedure developed by another laboratory, the forensic laboratory must first demonstrate reliability of the procedure in-house. Seven phases were designed to validate the use of the AmpFlSTR Profiler Plus PCR Amplification Kit, as well as the PE Applied Biosystems 310 Genetic Analyzer. This report summarizes the results obtained for each of the seven phases of the validation study which included the following evaluations: polymer, reproducibility, sensitivity, stutter, precision, mixtures and nonprobative casework. |
Databáze: |
MEDLINE |
Externí odkaz: |
|