Laser in situ keratomileusis for myopia and hyperopia using the Lasersight 200 laser in 300 consecutive eyes.

Autor: Reviglio VE; Fundación VER, Córdoba, Argentina., Bossana EL, Luna JD, Muiño JC, Juarez CP
Jazyk: angličtina
Zdroj: Journal of refractive surgery (Thorofare, N.J. : 1995) [J Refract Surg] 2000 Nov-Dec; Vol. 16 (6), pp. 716-23.
DOI: 10.3928/1081-597X-20001101-07
Abstrakt: Purpose: To evaluate effectiveness, safety, predictability, and short-term stability of laser in situ keratomileusis (LASIK) using the LaserSight Compac-200 Mini excimer laser with software version 9.0, for all refractive errors.
Methods: One hundred fifty consecutive patients (300 eyes) that received bilateral LASIK for myopia, hyperopia, and astigmatism were studied prospectively. A new 9.0 software version applying a modified nomogram that takes advantage of bilateral surgery was used. Follow-up at 6 months was available for 267 eyes (89%).
Results: Six months postoperatively, 131 eyes (96.32%) in the low to moderate myopia group (-1.00 to -5.99 D; n=136) had a spherical equivalent refraction within +/-1.00 D, and 123 eyes (90.44%) were within +/-0.50 D of emmetropia. In the high to extreme myopia group (-6.00 to -25.00 D; n=114), 97 eyes (87.08%) had a spherical equivalent refraction within +/-1.00 D and 78 eyes (68.42%) were within +/-0.50 D of emmetropia. In the hyperopia group (+1.00 to +6.00 D; n=50), 44 eyes (88%) had a postoperative spherical equivalent refraction within +/-1.00 D, and 31 eyes (62%) were within +/-0.50 D of emmetropia. Mean change in spherical equivalent refraction at 6 months was less than -0.50 D in the low to high myopia groups and -1.16 +/- 0.55 D in the extreme myopia group. At 6 months follow-up, uncorrected visual acuity was 20/20 or better in 73 eyes (54%) in the low to moderate myopia groups and 21 eyes (18%) in the high to extreme myopia groups. In the hyperopia group at 6 months follow-up, uncorrected visual acuity was 20/20 or better in 31 eyes (62%) and 20/40 or better in 41 eyes (82%). Only two eyes had a temporary loss of two or more lines of spectacle-corrected visual acuity due to corneal folds that were surgically treated. Six months after LASIK, no eye had lost any lines of best spectacle-corrected visual acuity in this series.
Conclusions: Our modified LASIK nomogram with the 9.0 software of the LaserSight 200 excimer laser (with a larger and smoother ablation pattern) resulted in safe and effective outcomes for the treatment of low to high myopia, astigmatism, and hyperopia.
Databáze: MEDLINE