Autor: |
Siderys H, Herod GT, Halbrook H, Pittman JN, Rubush JL, Kasebaker V, Berry GR Jr |
Jazyk: |
angličtina |
Zdroj: |
The Journal of thoracic and cardiovascular surgery [J Thorac Cardiovasc Surg] 1975 May; Vol. 69 (5), pp. 708-12. |
Abstrakt: |
A prospective clinical study involving more than 500 patients was designed to compare the Landé-Edwards membrane oxygenator and the Bentley bubble oxygenator. First, the importance of exposure of blood to the pericardium as the major source of hemolysis during open-heart surgery was confirmed. Because of this finding, we included in this study only those patients in whom the blood spilled into the pericardium was not returned to the pump. Under these circumstances, we found that hemolysis was relatively low in patients oxygenated with the membrane oxygenator. |