A practical gait analysis system using gyroscopes.

Autor: Tong K; Bioengineering Unit, Wolfson Centre, University of Strathclyde, Glasgow, UK., Granat MH
Jazyk: angličtina
Zdroj: Medical engineering & physics [Med Eng Phys] 1999 Mar; Vol. 21 (2), pp. 87-94.
DOI: 10.1016/s1350-4533(99)00030-2
Abstrakt: This study investigated the possibility of using uni-axial gyroscopes to develop a simple portable gait analysis system. Gyroscopes were attached on the skin surface of the shank and thigh segments and the angular velocity for each segment was recorded in each segment. Segment inclinations and knee angle were derived from segment angular velocities. The angular signals from a motion analysis system were used to evaluate the angular velocities and the derived signals from the gyroscopes. There was a good correlation between these signals. When performing a turn the signals of segment inclination and knee angle drifted. Two methods were used to solve this: automatically resetting the system to re-initialise the angle in each gait cycle, and high-pass filtering. They both successfully corrected this drift. A single gyroscope on the shank segment could provide information on segment inclination range, cadence, number of steps, and an estimation of stride length and walking speed.
Databáze: MEDLINE