Validity and reliability of total body volume and relative body fat mass from a 3-dimensional photonic body surface scanner.

Autor: Carolin Adler, Astrid Steinbrecher, Lina Jaeschke, Anja Mähler, Michael Boschmann, Stephanie Jeran, Tobias Pischon
Jazyk: angličtina
Rok vydání: 2017
Předmět:
Zdroj: PLoS ONE, Vol 12, Iss 7, p e0180201 (2017)
Druh dokumentu: article
ISSN: 1932-6203
DOI: 10.1371/journal.pone.0180201
Popis: Three-dimensional photonic body surface scanners (3DPS) feature a tool to estimate total body volume (BV) from 3D images of the human body, from which the relative body fat mass (%BF) can be calculated. However, information on validity and reliability of these measurements for application in epidemiological studies is limited.Validity was assessed among 32 participants (men, 50%) aged 20-58 years. BV and %BF were assessed using a 3DPS (VitusSmart XXL) and air displacement plethysmography (ADP) with a BOD POD® device using equations by Siri and Brozek. Three scans were obtained per participant (standard, relaxed, exhaled scan). Validity was evaluated based on the agreement of 3DPS with ADP using Bland Altman plots, correlation analysis and Wilcoxon signed ranks test for paired samples. Reliability was investigated in a separate sample of 18 participants (men, 67%) aged 25-66 years using intraclass correlation coefficients (ICC) based on two repeated 3DPS measurements four weeks apart.Mean BV and %BF were higher using 3DPS compared to ADP, (3DPS-ADP BV difference 1.1 ± 0.9 L, p