Autor: |
Tao Wu, Weidong Chen, Eric Fertein, Pascal Masselin, Xiaoming Gao, Weijun Zhang, Yingjian Wang, Johannes Koeth, Daniela Brückner, Xingdao He |
Jazyk: |
angličtina |
Rok vydání: |
2014 |
Předmět: |
|
Zdroj: |
Sensors, Vol 14, Iss 5, Pp 9027-9045 (2014) |
Druh dokumentu: |
article |
ISSN: |
1424-8220 |
DOI: |
10.3390/s140509027 |
Popis: |
A compact isotope ratio laser spectrometry (IRLS) instrument was developed for simultaneous measurements of the D/H, 18O/16O and 17O/16O isotope ratios in water by laser absorption spectroscopy at 2.73 μm. Special attention is paid to the spectral data processing and implementation of a Kalman adaptive filtering to improve the measurement precision. Reduction of up to 3-fold in standard deviation in isotope ratio determination was obtained by the use of a Fourier filtering to remove undulation structure from spectrum baseline. Application of Kalman filtering enables isotope ratio measurement at 1 s time intervals with a precision ( |
Databáze: |
Directory of Open Access Journals |
Externí odkaz: |
|