Analysis of the Influence of Different Settings of Scan Sequence Parameters on Vibration and Noise Generated in the Open-Air MRI Scanning Area

Autor: Jiří Přibil, Anna Přibilová, Ivan Frollo
Jazyk: angličtina
Rok vydání: 2019
Předmět:
Zdroj: Sensors, Vol 19, Iss 19, p 4198 (2019)
Druh dokumentu: article
ISSN: 1424-8220
DOI: 10.3390/s19194198
Popis: A system of gradient coils of the magnetic resonance imaging (MRI) device produces significant vibration and noise. Energetic relations of these phenomena are analyzed depending on MRI scan parameters (sequence type, repetition time (TR), echo time (TE), slice orientation, body weight). This issue should be investigated because of negative physiological and psychological effects on a person exposed to vibration and acoustic noise. We also measured the sound pressure level in the MRI scanning area and its vicinity in order to minimize these negative impacts, depending on intensity and time duration of exposition. From the recorded vibration and noise signals, the energy parameters were determined and statistically analyzed, and the obtained results were visually and numerically compared. Finally, subjective evaluation by a listening test method was used to analyze the influence of the generated MRI noise on the human psyche.
Databáze: Directory of Open Access Journals