Comparative Study of Different Drills for Bone Drilling: A Systematic Approach

Autor: Pazarci O, Torun Y, Ozturk A, Oztemur Z
Jazyk: angličtina
Rok vydání: 2020
Předmět:
Zdroj: Malaysian Orthopaedic Journal, Vol 14, Iss 2, Pp 83-89 (2020)
Druh dokumentu: article
ISSN: 1985-2533
2232-111X
DOI: 10.5704/MOJ.2007.016
Popis: INTRODUCTION: The performance of the drilling process depends on the characteristics of the drilling equipment and surgeon’s skill. To our knowledge, no research has focused on multi-parameter analysis of the dynamic behaviour of drills during the drilling process. This study aimed to characterise the physical changes and effects of different drills attached to a robotic arm during drilling of artificial bones in a standardised experimental setup. MATERIALS AND METHODS: Drilling processes using three brands of drills attached to a robotic arm were compared in terms of thrust force, vibration, noise level, speed deviation, and temperature. A standardised experimental setup was constructed, and measurement data were analysed statistically. Identical artificial bones were drilled 10 times with each drill. RESULTS: Thrust force measurements, which varied through the cortex and medulla, showed expressive differences for each drill for maximum and mean values (p
Databáze: Directory of Open Access Journals