Autor: |
Janis Kutz, Lars Liebermeister, Nico Vieweg, Konstantin Wenzel, Robert Kohlhaas, Mira Naftaly |
Jazyk: |
angličtina |
Rok vydání: |
2022 |
Předmět: |
|
Zdroj: |
Applied Sciences, Vol 12, Iss 16, p 8257 (2022) |
Druh dokumentu: |
article |
ISSN: |
2076-3417 |
DOI: |
10.3390/app12168257 |
Popis: |
We report calibration and performance tests of a terahertz fast-sweep optoelectronic frequency-domain spectrometer designed for industrial applications, aimed at quantifying its performance specifications and demonstrating its suitability for envisaged usage. The frequency scale is calibrated using atmospheric water vapour lines and a silicon wafer etalon; the amplitude linearity is verified using a set of silicon plates. Instrument performance is tested by measuring transmission properties of a variety of representative samples and comparing with a time-domain spectrometer and a frequency-domain spectrometer. |