The Sheared Central Venous Catheter?

Autor: Harihar V. Hegde, Vijay G. Yaliwal, Shyamsundar K. Joshi, P. Raghavendra Rao
Jazyk: angličtina
Rok vydání: 2011
Předmět:
Zdroj: Case Reports in Anesthesiology, Vol 2011 (2011)
Druh dokumentu: article
ISSN: 2090-6382
2090-6390
DOI: 10.1155/2011/379827
Popis: A fractured central venous catheter (CVC) with embolization of the distal fragment may lead to life-threatening complications. We had inserted a right subclavian CVC in a 68-year-old female which upon a follow-up chest X-ray appeared to have been sheared. A guidewire was inserted through the CVC until the J-tip was just beyond the tip of the CVC which were then withdrawn as a single assembly. We suspected that the tip of the guidewire might have been entrapped in the opening of the middle port, which upon withdrawal of the guidewire could have led to CVC folding upon itself and shearing.
Databáze: Directory of Open Access Journals