Cardiac transplant due to metal toxicity associated with hip arthroplasty

Autor: Sheldon Moniz, MBBS (UWA), Sean Hodgkinson, MBBS (UWA), Piers Yates, MBBS (Hons), BSc (Hons), MRCS (Eng), FRCS (Tr & Orth), FRACS (Ortho), FAOrthA
Jazyk: angličtina
Rok vydání: 2017
Předmět:
Zdroj: Arthroplasty Today, Vol 3, Iss 3, Pp 151-153 (2017)
Druh dokumentu: article
ISSN: 2352-3441
DOI: 10.1016/j.artd.2017.01.005
Popis: Concerns regarding metal-on-metal (MoM) bearing couples in total hip arthroplasty are well documented in the literature with cobalt (Co) and chromium (Cr) toxicity causing a range of both local and systemic adverse reactions. We describe the case of a patient undergoing cardiac transplantation as a direct result of Co and Cr toxicity following a MoM hip replacement. Poor implant positioning led to catastrophic wear generating abundant wear particles leading to Co and Cr toxicity, metallosis, bony destruction, elevated metal ion levels, and adverse biological responses. Systemic symptoms continued for 3 years following cardiac transplantation with resolution only after revision hip arthroplasty. There was no realization in the initial cardiac assessment and subsequent transplant workup that the hip replacement was the likely cause of the cardiac failure, and the hip replacement was not recognized as the cause until years after the heart transplant. This case highlights the need for clinicians to be aware of systemic MoM complications as well as the importance of positioning when using these prostheses.
Databáze: Directory of Open Access Journals