Validation of a wireless patch sensor to monitor mobility tested in both an experimental and a hospital setup: A cross-sectional study.

Autor: Niek Koenders, Joost P H Seeger, Teun van der Giessen, Ties J van den Hurk, Indy G M Smits, Anne M Tankink, Maria W G Nijhuis-van der Sanden, Thomas J Hoogeboom
Jazyk: angličtina
Rok vydání: 2018
Předmět:
Zdroj: PLoS ONE, Vol 13, Iss 10, p e0206304 (2018)
Druh dokumentu: article
ISSN: 1932-6203
DOI: 10.1371/journal.pone.0206304
Popis: PURPOSE:To assess the concurrent validity of a wireless patch sensor to monitor time lying, sitting/standing, and walking in an experimental and a hospital setup. METHODS:Healthy adults participated in two testing sessions: an experimental and real-world hospital setup. Data on time lying, sitting/standing, and walking was collected with the HealthPatch and concurrent video recordings. Validity was assessed in three ways: 1. test for mean differences between HealthPatch data and reference values; 2. Intraclass Correlation Coefficient analysis (ICC 3.1 agreement); and 3. test for mean differences between posture detection accuracies. RESULTS:Thirty-one males were included. Significant mean differences were found between HealthPatch data and reference values for sitting/standing (mean 14.4 minutes, reference: 12.0 minutes, p