Effect of MRI acquisition parameters on accuracy and precision of phase-contrast measurements in a small-lumen vessel phantom

Autor: Maria Correia de Verdier, Johan Berglund, Johan Wikström
Jazyk: angličtina
Rok vydání: 2024
Předmět:
Zdroj: European Radiology Experimental, Vol 8, Iss 1, Pp 1-11 (2024)
Druh dokumentu: article
ISSN: 2509-9280
DOI: 10.1186/s41747-024-00435-3
Popis: Abstract Background Phase-contrast magnetic resonance imaging (PC-MRI) quantifies blood flow and velocity noninvasively. Challenges arise in neurovascular disorders due to small vessels. We evaluated the impact of voxel size, number of signal averages (NSA), and velocity encoding (VENC) on PC-MRI measurement accuracy and precision in a small-lumen vessel phantom. Methods We constructed an in vitro model with a constant flow rate using a 2.2-mm inner diameter plastic tube. A reservoir with a weight scale and timer was used as standard reference. Gradient-echo T1 weighted PC-MRI sequence was performed on a 3-T scanner with varying voxel size (2.5, 5.0, 7.5 mm3), NSA (1, 2, 3), and VENC (200, 300, 400 cm/s). We repeated measurements nine times per setting, calculating mean flow rate, maximum velocity, and least detectable difference (LDD). Results PC-MRI flow measurements were higher than standard reference values (mean ranging from 7.3 to 9.5 mL/s compared with 6.6 mL/s). Decreased voxel size improved accuracy, reducing flow rate measurements from 9.5 to 7.3 mL/s. The LDD for flow rate and velocity varied between 1 and 5%. The LDD for flow rate decreased with increased voxel size and NSA (p = 0.033 and 0.042). The LDD for velocity decreased with increased voxel size (p