Comparing IDREAM as an Iterative Reconstruction Algorithm against In Filtered Back Projection in Computed Tomography

Autor: Eslam Maamoun
Jazyk: angličtina
Rok vydání: 2020
Předmět:
Zdroj: Iranian Journal of Medical Physics, Vol 17, Iss 3, Pp 170-174 (2020)
Druh dokumentu: article
ISSN: 2345-3672
DOI: 10.22038/ijmp.2019.41872.1613
Popis: Introduction: Recent studies of Computed Tomography (CT) conducted on patient dose reduction have recommended using an iterative reconstruction algorithm and mA (mili-Ampere) dose modulation. The current study aimed to evaluate Iterative Dose Reduction Algorithm (IDREAM) as an iterative reconstruction algorithm. Material and Methods: Two CT protocols (i.e., A: 120 KV /150 mA, FBP; B: 120KV/ (20-150) mAs, IDREAM) to scan water and acrylic phantoms. A number of 40 patients were assigned to two CT protocols (C: n=20, 120KV/160 ±10 mAs, FBP and D: n=20, 120 KV/ (30-150 mAs, IDREAM), the two groups (C and D) were then referred to abdomen and pelvis CT scan (Sinovision, insitum 16) with contrast. Image quality parameters, dose calculations were measured for all groups (i.e., A, B, C, and D). Results: Group B had a highly significant SNR with less significant noise (P0.05) in acrylic phantom, as compared to group A. CTDIvol (A: 13.94 mGy ; B: 6.91 mGy , PConclusion: The results of the present study were indicative of the feasibility of IDREAMas an iterative reconstruction algorithm.