[Delivery of desflurane by a Tec 6 vaporizer at a concentration superior to that posted on the regulating valve]
Autor: | D, Kayser, M, Khodja, G, Marguerite |
---|---|
Jazyk: | francouzština |
Rok vydání: | 1999 |
Předmět: | |
Zdroj: | Annales francaises d'anesthesie et de reanimation. 18(6) |
ISSN: | 0750-7658 |
Popis: | We report the case of a defective Tec 6 vaporizer (Datex-Ohmeda) which delivered higher concentrations of desflurane (end-tidal concentration: 3.5 vol%) than indicated by the vaporizer dial setting (2 vol%), at a fresh gas flow of 1 L.min-1. This condition was caused by a defective rotary control valve. Therefore it is essential to administer desflurane only in presence of a vapour analyser. |