[Delivery of desflurane by a Tec 6 vaporizer at a concentration superior to that posted on the regulating valve]

Autor: D, Kayser, M, Khodja, G, Marguerite
Jazyk: francouzština
Rok vydání: 1999
Předmět:
Zdroj: Annales francaises d'anesthesie et de reanimation. 18(6)
ISSN: 0750-7658
Popis: We report the case of a defective Tec 6 vaporizer (Datex-Ohmeda) which delivered higher concentrations of desflurane (end-tidal concentration: 3.5 vol%) than indicated by the vaporizer dial setting (2 vol%), at a fresh gas flow of 1 L.min-1. This condition was caused by a defective rotary control valve. Therefore it is essential to administer desflurane only in presence of a vapour analyser.