Autor: |
L P, Panych, C, Oesterle, G P, Zientara, J, Hennig |
Rok vydání: |
1996 |
Předmět: |
|
Zdroj: |
Magnetic resonance in medicine. 35(4) |
ISSN: |
0740-3194 |
Popis: |
In the paper, the results of a fast gradient-echo implementation of the singular value decomposition (SVD) encoding technique for dynamic imaging are presented. The method used is an adaptation with several critical modifications of a keyholetype approach previously proposed but not implemented. The method was tested by imaging the events following injection of a contrast agent into a phantom, producing a series of dynamic image updates. It is demonstrated that, for this type of application, the SVD encoding technique adequately follows dynamic changes with even a small number of encodes. The result is compared qualitatively to that obtained by standard Fourier-based keyhole imaging and is shown to provide improved spatial resolution of dynamic events when updating with the same number of encodes. |