Autor: |
Eduardo, Martínez-Morillo, Sandra, Elena-Pérez, David, Cembrero-Fuciños, María Francisca, García-Codesal, Teresa, Contreras-Sanfeliciano |
Rok vydání: |
2022 |
Předmět: |
|
Zdroj: |
Scandinavian journal of clinical and laboratory investigation. 82(5) |
ISSN: |
1502-7686 |
Popis: |
The verification of examination procedures is a responsibility for clinical laboratories in order to guarantee that their performance characteristics comply with the specifications obtained during the validation process and are congruent with the intended scope of the assay. The aim was to perform an evaluation of precision, bias, linearity, linear drift, sample carry-over, and comparability of 73 assays from Siemens Healthineers, by following the CLSI EP10-A3 guidelines. The verification was performed by measuring 72 biochemical parameters in quality control (QC) materials from Bio-Rad (except for IL6) with 73 assays installed on eight measuring systems (five Atellica |
Databáze: |
OpenAIRE |
Externí odkaz: |
|
Nepřihlášeným uživatelům se plný text nezobrazuje |
K zobrazení výsledku je třeba se přihlásit.
|