Management of complications of Sideris transcatheter devices for atrial septal defect closure

Autor: F A, Arabia, L J, Rosado, T R, Lloyd, G K, Sethi
Rok vydání: 1993
Předmět:
Zdroj: The Journal of thoracic and cardiovascular surgery. 106(5)
ISSN: 0022-5223
Popis: Various devices that can be inserted transvenously to close an ostium secundum atrial septal defect are undergoing clinical trials. Although these are safe and effective in most instances, they may occasionally dislodge or fail to "button" properly, causing migration and embolization. We report two cases in which the occluder and counteroccluder of the Sideris device for transvenous atrial septal defect occlusion (Custom Medical Devices, Amarillo, Tex.) failed to "button" appropriately, migrating in the right atrium in one patient and embolizing to the pulmonary artery in the second patient. An emergency operation was required to retrieve the device and repair the atrial septal defect.
Databáze: OpenAIRE