Compact multileaf collimator for conformal and intensity modulated fast neutron therapy: electromechanical design and validation

Autor: J B, Farr, R L, Maughan, M, Yudelev, E, Blosser, J, Brandon, T, Horste, J D, Forman
Rok vydání: 2006
Předmět:
Zdroj: Medical physics. 33(9)
ISSN: 0094-2405
Popis: The electromechanical properties of a 120-leaf, high-resolution, computer-controlled, fast neutron multileaf collimator (MLC) are presented. The MLC replaces an aging, manually operated multirod collimator. The MLC leaves project 5 mm in the isocentric plane perpendicular to the beam axis. A taper is included on the leaves matching beam divergence along one axis. The 5-mm leaf projection width is chosen to give high-resolution conformality across the entire field. The maximum field size provided is 30 x 30 cm2. To reduce the interleaf transmission a 0.254-mm blocking step is included. End-leaf steps totaling 0.762 mm are also provided allowing opposing leaves to close off within the primary radiation beam. The neutron MLC also includes individual 45 degrees and 60 degrees automated universal tungsten wedges. The automated high-resolution neutron collimation provides an increase in patient throughput capacity, enables a new modality, intensity modulated neutron therapy, and limits occupational radiation exposure by providing remote operation from a shielded console area.
Databáze: OpenAIRE