The objective of this study was to demonstrate the possibility to use deuterated compounds as internal standards for the quantitative analysis of morphine by gas chromatography with mass-selective detection for the purpose of doping control. The paper is focused on the problems associated with the use of deuterated morphine-D3 as the internal standard. Quantitative characteristics of the calibration dependence thus documented are presented along with uncertainty values obtained in the measurements with the use of deuterated morphine-D6. An approach to the assessment of method bias associated with the application of morphine-D6 as the deuterated internal standard is described. |