Autor: |
D, Hellingman, S C, Teixeira, M L, Donswijk, E J, Rijkhorst, L, Moliner, J, Alamo, C E, Loo, R A, Valdés Olmos, M P M, Stokkel |
Rok vydání: |
2016 |
Předmět: |
|
Zdroj: |
Revista espanola de medicina nuclear e imagen molecular. 36(3) |
ISSN: |
2253-8070 |
Popis: |
To assess the 3D geometric sampling accuracy of a new PET-guided system for breast cancer biopsy (BCB) from areas within the tumour with highIn the context of the European Union project MammoCare, a prototype semi-robotic stereotactic prototype BCB-device was incorporated into a dedicated high resolution PET-detector for breast imaging. The system consists of 2 stacked rings, each containing 12 plane detectors, forming a dodecagon with a 186mm aperture for 3D reconstruction (1mmNeedle positioning tests revealed an average accuracy of 0.5mm (range 0-1mm), 0.6mm (range 0-2mm), and 0.4mm (range 0-2mm) for the x/y/z-axes, respectively. Furthermore, the MammoCare system was able to visualize and locate small (10mm) regions with highAccuracy testing demonstrated high-precision of this semi-automatic 3D PET-guided system for breast cancer core needle biopsy. Its clinical feasibility evaluation in breast cancer patients scheduled for neo-adjuvant chemotherapy will follow. |