Retrieval and clinical analysis of distraction-based dual growing rod constructs for early-onset scoliosis
Autor: | Srinidhi Nagaraja, Behrooz A. Akbarnia, Pablo Bonangelino, Jeff Pawelek, Peter Sturm, John B. Emans, Maureen L. Dreher, Genevieve Hill, William M. Kane, Joshua Cockrum, Paul D. Sponseller |
---|---|
Rok vydání: | 2017 |
Předmět: |
Male
medicine.medical_specialty Adolescent Radiography Context (language use) Scoliosis 03 medical and health sciences 0302 clinical medicine Distraction Medicine Humans Orthopedics and Sports Medicine Orthopedic Procedures Registries Range of Motion Articular Child Orthodontics 030222 orthopedics Clinical pathology business.industry Infant medicine.disease Spine Surgery Child Preschool Equipment Failure Female Neurology (clinical) Implant Growing rod business Early onset scoliosis 030217 neurology & neurosurgery |
Zdroj: | The spine journal : official journal of the North American Spine Society. 17(10) |
ISSN: | 1878-1632 |
Popis: | Background Context Growing rod constructs are an important contribution for treating patients with early-onset scoliosis. These devices experience high failure rates, including rod fractures. Purpose The objective of this study was to identify the failure mechanism of retrieved growing rods, and to identify differences between patients with failed and intact constructs. Study Design/Setting Growing rod patients who had implant removal and were previously enrolled in a multicenter registry were eligible for this study. Patient Sample Forty dual-rod constructs were retrieved from 36 patients across four centers, and 34 of those constructs met the inclusion criteria. Eighteen constructs failed due to rod fracture. Sixteen intact constructs were removed due to final fusion (n=7), implant exchange (n=5), infection (n=2), or implant prominence (n=2). Outcome Measures Analyses of clinical registry data, radiographs, and retrievals were the outcome measures. Methods Retrievals were analyzed with microscopic imaging (optical and scanning electron microscopy) for areas of mechanical failure, damage, and corrosion. Failure analyses were conducted on the fracture surfaces to identify failure mechanism(s). Statistical analyses were performed to determine significant differences between the failed and intact groups. Results The failed rods fractured due to bending fatigue under flexion motion. Construct configuration and loading dictate high bending stresses at three distinct locations along the construct: (1) mid-construct, (2) adjacent to the tandem connector, or (3) adjacent to the distal anchor foundation. In addition, high torques used to insert set screws may create an initiation point for fatigue. Syndromic scoliosis, prior rod fractures, increase in patient weight, and rigid constructs consisting of tandem connectors and multiple crosslinks were associated with failure. Conclusion This is the first study to examine retrieved, failed growing rod implants across multiple centers. Our analysis found that rod fractures are due to bending fatigue, and that stress concentrations play an important role in rod fractures. Recommendations are made on surgical techniques, such as the use of torque-limiting wrenches or not exceeding the prescribed torques. Additional recommendations include frequent rod replacement in select patients during scheduled surgeries. |
Databáze: | OpenAIRE |
Externí odkaz: |