Popis: |
Bakeries add extra potassium bromate to the dough to make homogeneous, elastic, fluffy bread. Bromate causes renal damage and cancer. FAO/WHO stated that bromate residues shouldn't be in baked products. A potentiometric sensor's membrane recipe was optimized for sensitive and selective bromate assay. We planned a custom experimental design of 21 sensors that included numerical and categorical factors (NPPE: PVC, matrix%, membrane thickness, and ionophore type). We defined sensor performance outcomes (Nernstian slope, quantification limit, correlation coefficient, response time and selectivity), and each sensor's outcome was determined. The computer software developed a predictive model for each outcome and the desirability function suggested the optimum sensor recipe. The sensor achieved a slope of -63.54 mV/decade and detection limit of 2 × 10 |