Clinical evaluation of the Terry surgical keratometer

Autor: Richard P. Kratz, Bernard Davidson, Thomas R. Mazzocco, D. Michael Colvard
Rok vydání: 1980
Předmět:
Zdroj: American Intra-Ocular Implant Society Journal. 6:249-251
ISSN: 0146-2776
Popis: A study was designed to evaluate the usefulness of the Terry quantitative surgical keratometer as a tool to minimize surgically induced astigmatism. Two hundred and twenty-five phacoemulsifications with Shearing-style implants were performed, using identical technique. Seventy-five control cases were closed without the keratometer and 150 were closed with the keratometer. In the control group, 29.3% cases had corneal astigmatism of 3D or more, compared to 4.6% cases in the keratometer group (P.01). The total change in astigmatism was 2.64D in the control group and 0.91D in the keratometer group (P.02). We believe that these data demonstrate the usefulness of the Terry keratometer.
Databáze: OpenAIRE