Comparison of the Keeler Pulsair 2000 non-contact tonometer with Goldmann applanation

Autor: Gerald Naylor, Robert H Taylor, Anne Tan-Yee, John Deutsch, Alistair R. Fielder, J R Thompson, Merrick J. Moseley, Gary P. Misson
Rok vydání: 1993
Předmět:
Zdroj: Eye (London, England). 7
ISSN: 0950-222X
Popis: The Pulsair 2000 non-contact tonometer (Keeler Ltd, UK) is compared with the Goldmann applanation tonometer. Data from 80 eyes were acquired by four experienced observers. A linear regression analysis showed the relationship between the instruments to be: Pulsair = 0.66 + 0.95 Goldmann. Individual components of variation were analysed by analysis of variance which indicated a significant variation in the slope of the regression equation due to observers (p = 0.02) but not to the order in which topical anaesthesia was administered. Differences between two Pulsair instruments were of marginal significance (p = 0.07). The intercept of the regression equation was unaffected by any of the components of variation. Seventy-nine per cent of averaged intraocular pressure measurements obtained with the Pulsair 2000 fell on or within +/- 3 mmHg of those measured with the Goldmann tonometer. It is concluded that the Pulsair 2000 can provide clinically useful measurements of intraocular pressure.